Trios Error Codes

The information provided in this article is intended for use by licensed dental and healthcare professionals only, for informational purposes only, and does not constitute medical or dental advice or replace clinical judgment. The treating clinician bears sole responsibility for all diagnostic and therapeutic decisions made in connection with patient care.

Below you will find common error codes for the Trios Scanner and possible solutions to fix them.

Trios Error Codes

Case submision failures Case submission failed. Please contact Dandy support, or try again later. 
Post processing failures (Case submission) Post processing this order has encountered an issue. Please contact support for assistance in submitting this order. 
Post processing failures (Place order modal) Your scans could not finish processing. Return to review and try again, or contact Dandy support. 
Chat connection failures Failed to initiate chat with a lab tech 
Scanner connection errors Couldn't connect to the scanner 
Compare scans errors Failed to load the scan. Please choose another one. 
Patient case errors Something went wrong and we couldn't load this scan 
Multi-step continuation issuesNot deliveredWe cannot continue since the earlier order has not been delivered yet! 
Not Eligible (Crown)Cannot continue from an order that doesn't have a temporary crown, waxup, or matrix. 
Not Eligible (Denture)Cannot continue from an order that doesn't have a bite rim. 
Not FoundWe could not find that order! 
Submitting without scans Error submitting refabrication without scans: {{error}} 
Submitting refabrication errors Error submitting refabrication: {{error}} 
Re-use scans An error occurred when trying to copy one or more scans. 
Scanning stateGeneral errorsScanning error, Please check your scanner and try again. 
 WarmingScanner is warning 
Bite scan issuesAI bite alignment failuresAI-assisted bite alignment failed, Please align the bite manually or try re-scanning 
Bite alignment failuresBite alignment failed, Please rescan the bite or align it manually after scanning is complete. 
“Send to another lab” failures Post processing this order has encountered an issue. Please contact support for assistance in submitting this order. 
Diagnostic scan cloud sync failures Saving scan to the cloud failed. Please contact Dandy support, or try again later. 
AI Scan Review failuresGeneral review failuresWe encountered an error while reviewing your scans. We recommend retrying the check, and then double-checking your scans before placing your order. 
Hole analysis check failuresAnalysis failed: {{error}} 
Occlusal clearance check failuresOcclusal clearance check failed 
General review errorsWe encountered an error running this check. You can retry or continue without it. 
Gap detection check failuresScan quality check failed 
Denture gap hole analysis check failuresDenture gap check failed 
Incomplete scan check failuresMissing data check failed 
Margin check failuresMargin identification failed 
Undercut check failuresUndercut check failed 
Anterior clearance check failuresAnterior clearance check failed 

Calibration errors (Trios)


 

code 10errorHeadlineCalibration failed
 
errorSummaryIt looks like the mirror is giving uneven colors.
solutionStepsClean the glass lens on the black inner tube with a microfiber cloth. Ensure there are no cracks or scratches.
Clean the mirror tip. Ensure there are no smudges, cracks, or scratches.
Insert the Calibration Target into the Adapter with the gray side (labeled B) facing up toward the mirror in the tip and try calibrating again. If it doesn't work, try a new clean mirror tip. If the issue persists, contact support below.
code 11errorHeadlineCalibration failed
 
errorSummaryIt looks like the color balance is off.
solutionStepsClean the glass lens on the black inner tube with a microfiber cloth. Ensure there are no cracks or scratches.",
"Clean the mirror tip. Ensure there are no smudges, cracks, or scratches.
Insert the Calibration Target into the Adapter with the gray side (labeled B) facing up toward the mirror in the tip and try calibrating again. If it doesn't work, try a new clean mirror tip. If the issue persists, contact support below.
code 12errorHeadlineColor calibration mirror test failed
 
errorSummaryIt looks like the mirror may be obstructed.
solutionStepsClean the mirror tip. Ensure there are no smudges, cracks, or scratches.
Try calibrating again. If it doesn't work, try a new clean mirror tip. If the issue persists, contact support below.
code 1338errorHeadline3D Calibration failed
 
errorSummaryThe scanner is disconnected or not plugged in.
solutionStepsCheck that all of your cables are plugged in.
Ensure the scanner has a green baselight. If the baselight is off, the scanner is not receiving power.
Once the scanner baselight is green, try calibrating again. If the issue persists, contact support below.
code 15errorHeadline3D Calibration failed
 
errorSummaryThis may be caused by not calibrating your scanner for too long.
solutionStepsPlease contact support below to resolve this issue.
code 2errorHeadline3D Calibration failed
 
errorSummaryIs the black cylindrical calibration tip mounted on the scanner?
solutionStepsMount the black cylindrical calibration tip onto the scanner with no mirror or protective tip attached. There is a rubber stop, so as soon as you feel some resistance, you are all set.
Place the scanner on the base with the scanner button facing up toward the ceiling. Then, try calibrating again. If the issue persists, contact support below.
code 2000errorHeadline3D Calibration failed
 
errorSummaryThe scanner is disconnected.
solutionStepsCheck that all of your cables are plugged in.
Ensure the scanner has a green baselight. If the baselight is off, the scanner is not receiving power.
Once the scanner baselight is green, try calibrating again. If the issue persists, contact support below.
code 4errorHeadline3D Calibration failed
 
errorSummaryIt looks like your 3D calibration tip hasn't been mounted properly.
solutionStepsTry removing the black cylindrical calibration tip. Check that there are no marks or smudges on the glass lens inside the neck of the scanner. If there are, use a microfiber cloth to gently clean the glass lens, applying minimal pressure. Mount the black cylindrical calibration tip onto the scanner again. There is a rubber stop. As soon as you feel some resistance, you are all set. 
Place the scanner on the base with the scanner button facing up toward the ceiling. Then, try calibrating again. If the issue persists, contact support below.
code 5errorHeadline3D Calibration failed
 
errorSummaryIt looks like your 3D calibration tip hasn't been mounted properly.
solutionStepsClean the glass lens on the black inner tube with a microfiber cloth. Ensure there are no cracks or scratches.
Try calibrating again. If this step fails, it is possible to complete the 3D calibration step using the color calibration target on side B.
Try calibrating again. If the issue persists, contact support below.
code 7errorHeadlineColor calibration failed
 
errorSummaryIt looks like the Color Calibration Kit isn't mounted as expected.
solutionStepsMake sure the mirror tip is mounted snugly.
Insert the Calibration Target into the Adapter with the color side (labeled A) facing up toward the mirror in the tip and try calibrating again. If it doesn't work, try a new, clean mirror tip. If the issue persists, contact support below.
code 8errorHeadlineColor calibration failed
 
errorSummaryIt looks like the observed colors are off (A side).
solutionStepsCheck that there are no marks or smudges on the glass lens inside the neck of the scanner. If there are, use a microfiber cloth to gently clean the glass lens, applying minimal pressure.
Ensure there are no marks of smudges on the mirror inside the tip, if there are, attempt this step with a clean mirror tip. Make sure the mirror tip is mounted snugly.
Insert the Calibration Target into the Adapter with the color side (labeled A) facing up toward the mirror in the tip and try calibrating again. If the issue persists, contact support below.
code 9errorHeadlineColor calibration failed
 
errorSummaryIt looks like the Color Calibration Kit isn't mounted as expected.
solutionStepsMake sure the mirror tip is mounted snugly.
Insert the Calibration Target into the Adapter with the color side (labeled A) facing up toward the mirror in the tip and try calibrating again. If it doesn't work, try a new clean mirror tip. If the issue persists, contact support below.
code Minus 1errorHeadlineCalibration failed
 
errorSummarySomething went wrong.
solutionStepsCheck that all of your cables are plugged in.
Ensure the scanner has a green baselight. If the baselight is off, the scanner is not receiving power.
Once the scanner baselight is green, try calibrating again. If the issue persists, contact support below.
If you were on the color calibration step, make sure the appropriate Calibration Target is properly inserted into the adapter.
fallbackerrorHeadlineCalibration failed
 
errorSummarySomething went wrong

 

Was this article helpful?