The information provided in this article is intended for use by licensed dental and healthcare professionals only, for informational purposes only, and does not constitute medical or dental advice or replace clinical judgment. The treating clinician bears sole responsibility for all diagnostic and therapeutic decisions made in connection with patient care.

Below you will find common error codes for the Trios Scanner and possible solutions to fix them.

Trios Error Codes

Case submision failures   Case submission failed. Please contact Dandy support, or try again later.  
Post processing failures (Case submission)   Post processing this order has encountered an issue. Please contact support for assistance in submitting this order.  
Post processing failures (Place order modal)   Your scans could not finish processing. Return to review and try again, or contact Dandy support.  
Chat connection failures   Failed to initiate chat with a lab tech  
Scanner connection errors   Couldn't connect to the scanner  
Compare scans errors   Failed to load the scan. Please choose another one.  
Patient case errors   Something went wrong and we couldn't load this scan  
Multi-step continuation issues Not delivered We cannot continue since the earlier order has not been delivered yet!  
Not Eligible (Crown) Cannot continue from an order that doesn't have a temporary crown, waxup, or matrix.  
Not Eligible (Denture) Cannot continue from an order that doesn't have a bite rim.  
Not Found We could not find that order!  
Submitting without scans   Error submitting refabrication without scans: {{error}}  
Submitting refabrication errors   Error submitting refabrication: {{error}}  
Re-use scans   An error occurred when trying to copy one or more scans.  
Scanning state General errors Scanning error, Please check your scanner and try again.  
  Warming Scanner is warning  
Bite scan issues AI bite alignment failures AI-assisted bite alignment failed, Please align the bite manually or try re-scanning  
Bite alignment failures Bite alignment failed, Please rescan the bite or align it manually after scanning is complete.  
“Send to another lab” failures   Post processing this order has encountered an issue. Please contact support for assistance in submitting this order.  
Diagnostic scan cloud sync failures   Saving scan to the cloud failed. Please contact Dandy support, or try again later.  
AI Scan Review failures General review failures We encountered an error while reviewing your scans. We recommend retrying the check, and then double-checking your scans before placing your order.  
Hole analysis check failures Analysis failed: {{error}}  
Occlusal clearance check failures Occlusal clearance check failed  
General review errors We encountered an error running this check. You can retry or continue without it.  
Gap detection check failures Scan quality check failed  
Denture gap hole analysis check failures Denture gap check failed  
Incomplete scan check failures Missing data check failed  
Margin check failures Margin identification failed  
Undercut check failures Undercut check failed  
Anterior clearance check failures Anterior clearance check failed  

Calibration errors (Trios)


 

code 10 errorHeadline Calibration failed
 
errorSummary It looks like the mirror is giving uneven colours.
solutionSteps Clean the glass lens on the black inner tube with a microfiber cloth. Ensure there are no cracks or scratches.
Clean the mirror tip. Ensure there are no smudges, cracks, or scratches.
Insert the Calibration Target into the Adapter with the gray side (labelled B) facing up toward the mirror in the tip and try calibrating again. If it doesn't work, try a new clean mirror tip. If the issue persists, contact support below.
code 11 errorHeadline Calibration failed
 
errorSummary It looks like the colour balance is off.
solutionSteps Clean the glass lens on the black inner tube with a microfiber cloth. Ensure there are no cracks or scratches.",
"Clean the mirror tip. Ensure there are no smudges, cracks, or scratches.
Insert the Calibration Target into the Adapter with the gray side (labelled B) facing up toward the mirror in the tip and try calibrating again. If it doesn't work, try a new clean mirror tip. If the issue persists, contact support below.
code 12 errorHeadline Colour calibration mirror test failed
 
errorSummary It looks like the mirror may be obstructed.
solutionSteps Clean the mirror tip. Ensure there are no smudges, cracks, or scratches.
Try calibrating again. If it doesn't work, try a new clean mirror tip. If the issue persists, contact support below.
code 1338 errorHeadline 3D Calibration failed
 
errorSummary The scanner is disconnected or not plugged in.
solutionSteps Check that all of your cables are plugged in.
Ensure the scanner has a green baselight. If the baselight is off, the scanner is not receiving power.
Once the scanner baselight is green, try calibrating again. If the issue persists, contact support below.
code 15 errorHeadline 3D Calibration failed
 
errorSummary This may be caused by not calibrating your scanner for too long.
solutionSteps Please contact support below to resolve this issue.
code 2 errorHeadline 3D Calibration failed
 
errorSummary Is the black cylindrical calibration tip mounted on the scanner?
solutionSteps Mount the black cylindrical calibration tip onto the scanner with no mirror or protective tip attached. There is a rubber stop, so as soon as you feel some resistance, you are all set.
Place the scanner on the base with the scanner button facing up toward the ceiling. Then, try calibrating again. If the issue persists, contact support below.
code 2000 errorHeadline 3D Calibration failed
 
errorSummary The scanner is disconnected.
solutionSteps Check that all of your cables are plugged in.
Ensure the scanner has a green baselight. If the baselight is off, the scanner is not receiving power.
Once the scanner baselight is green, try calibrating again. If the issue persists, contact support below.
code 4 errorHeadline 3D Calibration failed
 
errorSummary It looks like your 3D calibration tip hasn't been mounted properly.
solutionSteps Try removing the black cylindrical calibration tip. Check that there are no marks or smudges on the glass lens inside the neck of the scanner. If there are, use a microfiber cloth to gently clean the glass lens, applying minimal pressure. Mount the black cylindrical calibration tip onto the scanner again. There is a rubber stop. As soon as you feel some resistance, you are all set. 
Place the scanner on the base with the scanner button facing up toward the ceiling. Then, try calibrating again. If the issue persists, contact support below.
code 5 errorHeadline 3D Calibration failed
 
errorSummary It looks like your 3D calibration tip hasn't been mounted properly.
solutionSteps Clean the glass lens on the black inner tube with a microfiber cloth. Ensure there are no cracks or scratches.
Try calibrating again. If this step fails, it is possible to complete the 3D calibration step using the colour calibration target on side B.
Try calibrating again. If the issue persists, contact support below.
code 7 errorHeadline Colour calibration failed
 
errorSummary It looks like the Colour Calibration Kit isn't mounted as expected.
solutionSteps Make sure the mirror tip is mounted snugly.
Insert the Calibration Target into the Adapter with the colour side (labelled A) facing up toward the mirror in the tip and try calibrating again. If it doesn't work, try a new, clean mirror tip. If the issue persists, contact support below.
code 8 errorHeadline Colour calibration failed
 
errorSummary It looks like the observed colours are off (A side).
solutionSteps Check that there are no marks or smudges on the glass lens inside the neck of the scanner. If there are, use a microfiber cloth to gently clean the glass lens, applying minimal pressure.
Ensure there are no marks of smudges on the mirror inside the tip, if there are, attempt this step with a clean mirror tip. Make sure the mirror tip is mounted snugly.
Insert the Calibration Target into the Adapter with the colour side (labelled A) facing up toward the mirror in the tip and try calibrating again. If the issue persists, contact support below.
code 9 errorHeadline Colour calibration failed
 
errorSummary It looks like the Colour Calibration Kit isn't mounted as expected.
solutionSteps Make sure the mirror tip is mounted snugly.
Insert the Calibration Target into the Adapter with the colour side (labelled A) facing up toward the mirror in the tip and try calibrating again. If it doesn't work, try a new clean mirror tip. If the issue persists, contact support below.
code Minus 1 errorHeadline Calibration failed
 
errorSummary Something went wrong.
solutionSteps Check that all of your cables are plugged in.
Ensure the scanner has a green baselight. If the baselight is off, the scanner is not receiving power.
Once the scanner baselight is green, try calibrating again. If the issue persists, contact support below.
If you were on the colour calibration step, make sure the appropriate Calibration Target is properly inserted into the adapter.
fallback errorHeadline Calibration failed
 
errorSummary Something went wrong

 

Was this article helpful?