Dandy Vision Scanner Sterilization

Tip Sterilization

  • Remove the scanning tip from the scanner
    • Replace the scanning tip with the protective tip
    • Wipe the base with disinfecting wipes, being careful to avoid the air vents and the electric outlet of the scanner. Set aside.
      • Want: cleaning/disinfectant wipes containing Isopropanol (17.2%) and Ammonium Chloride.
  • Wash the scanning tip(s) with soap and warm water for 2 minutes.
  • Dry the tip(s)
  • Place a 4x4 gauze on the mirror(s) of the tip(s) before placing them in the autoclave bag(s)
    • It protects the mirror during sterilization and stays on throughout.
  • Autoclave at 273.2 degrees Fahrenheit (134 degrees Celsius) for 3-6 minutes to sterilize.
    • If your machine has a plastic setting, do not use this setting.
  • If visible smudges or debris are present on the scanner lens, use a microfiber cloth to wipe it clean.
    • It prevents scratching and deposition of filament residue on the lens