Alliedstar Scanner Sterilization

Tip Sterilization

  • Remove the scanning tip from the scanner
    • Replace the scanning tip w/ the protective tip
    • Wipe the base with disinfecting wipes - careful to avoid the air vents and electric outlet of the scanner. Set aside.
  • Wash the scanning tip(s) with soap and warm water for 2 minutes.
  • Dry the tip(s)
  • Place a 4x4 gauze on the mirror(s) of the tip(s) before placing in the autoclave bag(s)
  • Autoclave at 273.2 degrees fahrenheit (134 degrees celsius) for 3-6 minutes to sterilize
    • If your machine has a plastic setting, do not use this setting.