Instructions for cleaning and sterilizing the scanner components, including mirror tips, scanner lenses, and hand pieces.
Materials Needed
- Paper sterilization pouches
- Autoclave
- Wavicide-01 solution
- Lint-free, non-abrasive cloth
- Cotton-tipped swab
- Denatured alcohol - Ethanol or Propanol - (60–70% Alc/Vol)
Sterilizing Mirror Tips
- Place each scanner tip into a paper sterilization pouch, ensuring a tight seal
- Use an autoclave to sterilize the tips under one of the following conditions:
- 121℃ (249.8℉) for 30 minutes, with a subsequent 15-minute drying period
- 135℃ (275℉) for 10 minutes, followed by 30 minutes of drying
- Scanner tips are durable for up to 150 sterilization cycles. Follow the disposal guide for proper discarding
Cleaning and Disinfecting Scanner Tips
- Submerge the tip in Wavicide-01 for 45 to 60 minutes. Follow with a thorough rinse
-
- Note: Refer to the Wavicide-01 solution instruction manual for proper use
- Pat the tip dry with a sterilized, non-abrasive cloth
Caring for the Scanner Lens:
- Wipe the lens with a cloth or cotton-tipped swab moistened with alcohol. Ensure the use of ethanol or propanol to avoid staining
- Use a dry, lint-free cloth to remove any moisture
Maintaining the Scanner Hand Piece:
- Clean the filter using an alcohol-soaked cloth with the device off and disconnected. Cover filter when completed
- Note: Do not wipe with your hands or use excessive force.
- Apply denatured alcohol to a soft, lint-free, nonabrasive cloth and wipe the scanner's surface. Dry thoroughly