Enclosed Scan Tip
The Dandy Trios Wireless Scanner has a new 'Enclosed scan tip'. The new closed scan tip with scratch-free sapphire glass provides a microbial barrier:
Dandy Trios Wireless Wand
Single-use ultra-thin sleeves minimize the risk of cross-contamination:
- It is recommended to use the TRIOS Sleeve to cover the TRIOS scanner during patient examination.
- The TRIOS Sleeve must only be used once to avoid cross-contamination.
Cleaning the Wand
- Wearing gloves, remove TRIOS Sleeve and TRIOS scanner tip from TRIOS scanner body.
- Wipe the scanner body and mirror housing with any cleaning/disinfectant wipes containing Isopropanol (17.2%) and Ammonium Chloride (for example, CaviWipes).
- While still wet from the CaviWipe, use a dry soft bristled brush to clean around the gasket area (where the TRIOS scanner body and mirror housing connect).
- Wipe the surface of the TRIOS scanner body and mirror housing with a second wipe.
- Take a third wipe and wet the scanner body and mirror housing for minimum of 120 seconds.
- 120 seconds is the minimal time for the solution in the wipes to be in contact with the surface to kill the bacteria. It can be longer, but not less.
- Once done, place the protection tip on the scanner.
Cleaning and disinfecting TRIOS Pod (Scanner Base)
- Wearing gloves, remove the rubber insert from the pod (holder) by pressing down on the middle using a thumb and gripping the outer edge with the index finger:
- Using the cleaning/disinfectant wipes containing Isopropanol (17.2%) and Ammonium Chloride (for example, CaviWipes), clean the surface of the top rubber insert to remove any gross contaminants.
- With second wipe clean the surface of the metal holder to remove any gross contaminants.
- With third wipe wet the surface of the rubber insert for minimum 120 seconds to disinfect.
- With fourth wipe wet the surface of the metal part for minimum 120 seconds to disinfect.
- 120 seconds is the minimal time for the solution in the wipes to be in contact with the surface to kill the bacteria. It can be longer, but not less.
- Allow both parts to dry and assemble them back together.
Cleaning the Scanner Tip
- Single-use tips can be discarded after use
- Autoclavable Scanner Tips:
- Wearing gloves, clean the TRIOS scanner tip manually using a detergent diluted according to the manufacturer’s instructions. Apply the detergent directly on a soft-bristle brush. Wash the TRIOS scanner tip under running lukewarm (30-40 °C or 86-104 °F) utility water.
- Inspect the TRIOS scanner tip window after cleaning. If the window appears with stains, smudged or with a milky haze, repeat this step.
- Rinse with lukewarm (30-40 °C or 86-104 °F) tap water.
- Dry the TRIOS scanner tip carefully with a non-woven swab (for example, Abena non-woven swabs).
Sterilizing the Scanner Tip
- Single-use tips can be discarded after use
- Autoclavable Scanner Tips (once clean):
- Put the TRIOS scanner tip in a sterilization pouch. The pouch should be sealed. Use either a self-adhesive or heat-sealed pouch.
- Sterilize the pouched TRIOS scanner tip in a Dynamic-Air-Removal Steam Sterilizer (Class B autoclave):
- Autoclave the TRIOS scanner tip at 132 °C (269.6 °F) with a cycle of 4 minutes exposure time and 30 minutes dry time.
- The TRIOS scanner tip can be re-sterilized up to 100 times and must be disposed of afterwards.
- New TRIOS scanner tips must be cleaned and sterilized before using them for the first time.